Door opening
A single
lock located at the bottom right hand corner closes the door. The security key has a raised tab and this
should be aligned with the slot in the lock in order to open the door. The key can then be turned 90°
counter-clockwise.
Changing Chart
In order
to change the chart the centre hub must be removed. Take hold of the centre hub and turn the sprung, knurled centre
cap by rotating it counter-clockwise, until it releases. The chart can then be
slid off the instrument. Replace with a
new chart by sliding the chart under the retaining fold at the top of the
instrument and position over the centre hole. To replace the sprung centre
screw, re-align it with the hole, push the screw down and turn until it has
tightened (it should only be finger tight).
Aligning the chart with the pen – Press and hold the button located
on the bottom left of the instrument until the chart rotates to the correct
position.
Calibration
The
Mini-Chart recorder is pre-calibrated and should require no adjustment;
however, the pen position can be altered by holding the pen arm with finger and
thumb near to the spindle and gently turning the M3 lock nut on top of the
spindle with a nut-spinner. This can be
repeated until the pen nib indicates the required temperature.
Breakdown
In the
event of incorrect operation contact the manufacturer or the supplier for
advice. Dismantling will invalidate the warranty.
Abacus instruments, 13C Old Bridge Way, Shefford, Bedfordshire
SG17 5HQ, England, U.K., T: - 01462 638200, Fax: - 01462 638201
Website: - www.abacusinstruments.com
4
Instructions Manual - Mini-chart

For remote bulb
models
Carefully
uncoil the capillary and locate the sensor in the required position for optimum
monitoring.
Wall mounted For wall mounting remove the chart and use the top and bottom
holes on the rear of the instrument to fix to the wall with two suitably sized
screws. See diagram

1
Panel
mounted The panel mounted units have two
integral clips which enable them to be easily mounted into a DIN size cut-out
directly from the front. Insert the
instrument into the cut-out and turn the door lock counter-clockwise by 90°C to
open the door. Once the chart is
removed the two large screw heads in the centre of the top and bottom edges
should be gently tightened clockwise with a screwdriver to rotate and to draw
the two securing lugs up towards the back of the panel. The Mini-Chart is now mounted securely. See diagram.

Battery powered
The
battery is pre fitted and should not be removed. The battery is designed to
power the unit for up to 5 years. The battery should be replaced by a 3V
Lithium battery (positive is indicated on the label covering the battery
department. To check the battery press
and hold the button located on the bottom left of the unit, the chart should
rotate one full revolution.

Mains powered
The Mini-chart should be connected to a suitable power supply with the lead provided. When connected a red LED in the top left hand corner will be lit. If the LED fails to light then check the supply.
3
Changing the pen
Wait until the unit is
within its calibrated operating temperature range. Remove the pen from its packaging (remove & retain the tip
seal) and gently moving the pen arm lifter to the right the pen arm will rise,
slide the pen onto the pen arm taking care that the tip points toward the
chart. Once the pen is located correctly move the pen arm lifter back to the
original position. Undue force should
be avoided as the pen arm is sprung to ensure optimum contact pressure with the
chart. If at any time the pen arm is
bent so that insufficient force is being applied to make a trace on the paper
then the nib should be removed and light pressure exerted on the arm as near as
possible to the spindle until the other end of the pen arm is resting about 5mm
from the paper. The pen nib can then be
attached once again.
![]()
![]()
![]()
![]()

![]()